Número de pieza
Desmilitarización
Vigencia
UOM
NIIN
Número de pieza:
90-0978p
Información de la pieza
Blood gas analyzer 120/230V 50/60 hz ac:
NSN
Número de acciones nacionales:
TXT
Descripción:
Blood Gas Analyzer
Código de nombre del artículo:
38049
MRC:
A measurement from the bottom to the top of an object,in distinction from depth.ABRN
Height:
8.400 inches
MRC:
Indicates the type of current whether alternating,direct,or both.ACDC
Current Type:
Ac
MRC:
The number of complete cyclic changes,per unit of time,for which an item is rated.ACZB
Frequency Rating:
50.0 hertz or 60.0 hertz
MRC:
Indicates the type of test for which the item is designed.AQXY
Test Type For Which Designed:
Blood gases and electrolytes; HCO3; hct base excess and O2 sat
MRC:
Those features,not otherwise specified,which may be required for proper functioning of the item.CBBL
Features Provided:
Digital display screen and integral printer and self-calibration
MRC:
The method by which an instrument performs the actual assay(s) of a sample/specimen.CYFV
Method Of Analysis:
Polarographic and potentiometric/ion selective electrode
MRC:
The built in testing range for each type of test/assay that the analytical instrument is capable of performing.CYFY
Analysis Test/measurement Range:
PO2 0-600 mmhg ph 6.8 to 7.8; PCO2 8 to 99 mmhg k 0.1 to 9.9 mm/l na 100 to 200 mm/l ca 0.25 to 4.25 mm/l hct 10 to 75 pct HCO3 3.0 to 60 mm/l TCO2 3.0 to 60 mm/l base excess -30 to +30 mm/l O2 sat 30 to 99 pct
MRC:
The amount,mass,volume,etc.,of a sample/specimen needed by the item to run a complete assay/test.CYFZ
Test Sample/specimen Size:
0.5 ml
MRC:
Indicates the type of sample/specimen the item is capable of ascertaining qualitative and/or quantitative assay determinations.CYHM
Test Sample/specimen Type:
Whole blood
MRC:
A measurement between specified points on an item, in distinction from height.DPTH
Depth:
18.300 inches
MRC:
The total electrical voltage.ELEC
Voltage In Volts:
120.0 and 230.0
MRC:
Those unusual or unique characteristics or qualities of an item not covered in the other requirements and which are determined to be essential for identification.FEAT
Special Features:
Utilizes a disposable cartridge/test pack that contains all electrodes calibrating and flush solutions and gases plus a waste container for processed blood calibrating, and qc solutions The cartridge/test pack shall eliminate the need for gas tanks external solutions and electrode maintenance. Shall be pre-programed to automatically perform two-point calibration every hour and one-point calibration every 90 seconds Capable of automatic aspiration of the sample and to process display and print the results in less than 2 minutes Shall be suitable for the field environment.
MRC:
A measurement of volume taken by multiplying the length by the width by the height of the unit of issue and rendered in cubic units.SUCB
Unit Of Issue Cube:
1.00 cubic feet
MRC:
The measured weight of the actual container(s) or supporting device(s) which is in direct contact with the item and its contents.SUWT
Unit Of Issue Weight:
27.00 pounds
MRC:
A measurement taken at right angles to the length of an item,in distinction from thickness.WDTH
Width:
9.100 inches