Número de pieza 80063UN01

Intravenous Infusion Pump

Precios y disponibilidad

Envíe este formulario para conocer los precios actuales y la disponibilidad de esta NSN.
Número de pieza:
Cantidad:
Dirección de correo electrónico:
Teléfono:
Referencia:
Compañía:
Nombre:

No compartimos ni vendemos su información a nadie.
Privacidad | Letra chica

All major credit cards accepted as well as gov. p-cards
Número de pieza
Desmilitarización
Vigencia
UOM
NIIN
Número de pieza:
80063un01
Desmilitarización:
No
Vigencia:
Unidad de medida:
NIIN:
015607883

Información de la pieza
Alaris gp volumetric pump

NSN
Número de acciones nacionales:
6515-01-560-7883 6515015607883
TXT
Descripción:
Intravenous Infusion Pump
INC
INC
Código de nombre del artículo:
45742
MRC:
The distance measured in a straight line from the bottom to the top of an item.ABKW
Overall Height:
225.0 millimeters
MRC:
An overall measurement taken at right angles to the length of an item,in distinction from thickness.ABMK
Overall Width:
148.0 millimeters
MRC:
Indicates the type of current whether alternating,direct,or both.ACDC
Current Type:
Ac
MRC:
An indication of whether or not the item is portable.APCB
Portability:
Portable
MRC:
The total electrical voltage.ELEC
Voltage In Volts:
9.6
MRC:
Those unusual or unique characteristics or qualities of an item not covered in the other requirements and which are determined to be essential for identification.FEAT
Special Features:
For acute and sub-acute applications. Infusion rate range: 1-1200 ml/h Vtbi range primary: to 10000 ml Vi range: 1 ml to 50000 ml Volumetric accuracy: plus/minus 5 percent () Mechanism: four finger displacement peristaltic. Display: mono (240 x 240) pixels Free flow protection (alaris safety clamp) Kvo rates: 5 ml/h or set infusion rate if lower Occlusion setting: 3 levels from L2, L5, L8 (factory default L5). Battery type: rechargable nimh 9.6 v 2.4 ah Battery run time: operates for 4 hr from full charge at 125 ml/h Recharge time: 2 1/2 hrs from discharge to 95 percent charge Electrical protection: class i type cf (defibrillation proof) Ac power requirements: 100-230 vac 50-60HZ () Dimensions: 148M (w) x 225MM(H) x 148MM(D) Weight: 2.5KG.
MRC:
The cycles per second /hertz/ of the alternating current.FREQ
Frequency In Hertz:
50.0 or 60.0

Similar Intravenous Infusion Pumps

« Medical and Surgical Instruments, Equipment and Supplies Catálogo
Comparar ahora»
Claro | Esconder